The ARS PS-L-UHV Probe Station uses Liquid Helium or Liquid Nitrogen as the cooling agent to reach cryogenic temperatures of 4 K or 77 K depending upon the cryogen used.
The core of the system is the same for both the Liquid and Closed Cycle Probe Stations, which enables the Flow Cryostat to be removed from this system and replaced with a Closed Cycle (cryogen-free) configuration at a later point, if your needs change.
ARS manufactures both the Helitran Helium transfer system and the probe station. This integrated approach ensures consistent performance and also facilitates diagnostics and service of the integrated system.
Cooling Technology | ||
---|---|---|
LT-3 | Liquid Flow Cryostat | |
Liquid Cryogen Usage | Liquid Helium or Liquid Nitrogen | |
Temperature* | ||
Liquid Helium | ~3.5 K - 400 K* | |
Liquid Nitrogen | ~77 K - 400 K* | |
Stability | 0.1 K | |
* 500K and 800K options also available | ||
Cooldown Time | ||
~ 4.5 K | ~ 10 minutes | |
Sample Holders* Custom options also available | ||
Grounded | 2.25” Diameter (standard) | |
Load Lock – Grounded | 1” Diameter: Custom Sizes available | |
Electrically Isolated | 2.25” Diameter (Optional) | |
Coaxial (Biased) | 2.25” Diameter (Optional) Includes BNC Feedthrough with Coaxial Cable to sample stage | |
Triaxial (Biased with Guard) | 2.25” Diameter (Optional) Includes Triaxial Feedthrough with Coaxial or Triaxial wire. | |
Radiation Shield | ||
Material | Nickel Plated OFHC Copper | |
Diameter | 8” (203 mm) | |
Ports | 6 (standard) Empty ports have removable blanks installed | |
Lid | Removable Lid with Sapphire Cold Window | |
Mounted On | Standoffs for ideal thermal gradient | |
Optical Access | ||
Window Mat. | Sapphire | |
Window Diam. | 2.5” (63 mm) | |
Window Clear View | 2.25” (57 mm) |
Vacuum Shroud | ||
Material | Welded Conflats, Stainless Steel | |
Diameter | 11.97” (304 mm) | |
Probe Ports | 4 (standard): up to 6 - 8 available | |
Spare Accessory Ports | (2) UHV Flanges installed with blanks (1) 8” Diameter UHV Pump out Port |
|
Lid | Conflat Flanged Lid with Quartz Window | |
Mounted On | 31” x 31” Powder Coated Aluminum Plate | |
Optical Access | ||
Window Material | High Purity Quartz | |
Window Diam. | 2.75” (69 mm) | |
Window Clear View | 2.37” (60 mm) | |
Probe Arms | ||
---|---|---|
GSG Microwave | Tungsten with Coaxial or Triaxial Cable (custom tips available) |
|
DC/Low Frequency | Contact ARS for Custom Microwave Probe Solutions | |
Fiber Optic | Contact ARS for Custom Fiber-Optic Probe Solutions | |
Vibration Isolated Table | ||
Vibration Level | < 1 µm (With Cryocooler Running) | |
Table Top Height | 927 mm (36 in.) | |
Table Top Width | 927 mm ( 36 in.) | |
7:1 GigE Microscope with Zoom Lens* | ||
Field Of View | 4.2 - 0.61 mm | |
Working distance | 89 mm | |
Numerical Aperture (N.A.) | 0.024 - 0.08 | |
Highest Resolution | 1 µm | |
Magnification | Final magnification depends on the monitor size | |
Image Size | 1280 x 1024 pixels | |
Camera Output | USB-2.0 | |
Frame rate | 25 fps | |